S160 CO2/RH analyzer
Date:2024-12-19 08:26:36

image.png


Equipment introduction

S160co2/RH analyzer is a flow-through analyzer, the device will be based on a capacitive RH sensor and infrared CO that responds to quickly2The sensor is integrated into a temperature-controllable, rugged and durable weather protection box.

The CO2 analyzer has a measurement range of 0-2000ppm and a resolution of 1ppm.Extended range: 0-5000ppm CO2 (optional).

The RH sensor (range 0-100%) is associated with the temperature sensor, so the measured RH value is based on the temperature value of the measuring instrument.

The S160 has an analog output of CO2, RH, temperature and flow (if a flow monitor is configured).It can be used with Qubit Systems' analog digital data interface and software, or with the user's own data acquisition system.The S160 can also be equipped with a built-in air pump and flow monitor (G160-PF).


Application areas

• Animal respiration assay

•Insect Respiratory Assay

•Environmental monitoring

•Plant breathing

•Soil sample breathing

•other


Technical parameters


Relative Humidity (RH) Sensor

ComeThermoset polymer capacitor sensor

ComeLinear voltage output RH%

ComeLow power design

ComeAccuracy: ±2%RH, 0-100%RH, no condensation at 25°C

ComeLinearity: Typical ±0.5%RH

ComeError: ±1.2% of the maximum relative humidity range

ComeRepeatability: ±0.5%RH

ComeResponse time: 15 s in slow moving air at 25°C

ComeStability: Within 5 years, ±1%RH under 50%RH

ComeRange: 0-100%RH, no condensation

ComeOperating temperature range: 5°C-60°C

ComeTemperature compensation: Real RH = sensor RH/(1.0546-0.00216 T) T, unit °C

(Real RH=Sensor RH/(1.093-0.0012T)T, unit in °F)

ComeChemical corrosion resistance

Come0%RH V output = 0.200V

Come100%RH V output = 4.800V

ComeSlope = 21.739, offset = -4.3478

Temperature sensor

ComeNTC chip thermistor,Nonlinear response

ComeSteinhart equation in software provides degrees Celsius readings:

ComeAccuracy: +/-0.2C

ComeRange: 5°C to 60°C

Carbon dioxide analyzer T

ComeWorking principle: Non-dispersive infrared sensor

ComeGas sampling mode: flowing air flow, sealed chamber

ComeMaximum gas flow rate through the analyzer: 650 mL/min

ComeMeasurement range: (LCD) -0–1999 ppm

ComeResolution: ±1 ppm

ComeRepeatability (at stable AP and temperature): better than ±1 ppm

ComeMaximum drift (year): ±100 ppm

ComeResponse time (@250 mL/min; 95% of final value): about 25 seconds

ComeWarm-up time (@22oC): about 5 minutes

ComeloseOutput (linear): 0–5 VDC at 0–2000 ppm

ComeCalibration adjustment: Zero point and range

ComeOperating temperature range: 0 to 50oC

ComeStorage temperature range: -40 to 70oC

ComeOperating pressure range: ±1.5% local average pressure

ComeHumidity range: 5-95%RH, no condensation (dry airflow is recommended)

ComePressure dependence: +0.19% reading/mm Hg

ComePower supply: 12 VDC, via 120 VAC/60 Hz adapter

ComeCurrent: average 200 mA, peak 450 mA

ComeDimensions: Height x Width x Length = 17 x 31 x 39 cm

ComeWeight: 4.7kg