iTOXcontrol water quality comprehensive toxicity online monitoring system
Date:2017-05-05 13:56:33

Product Introduction

The latest generation of bio-integrated toxicity online meter developed by the Dutch microLAN company is named iTOXcontrol, which has the following characteristics and functions:

UA fully automatic culture device with bioluminescent bacteria is integrated into the instrument;

UThe data and related information of the instrument can be uploaded to an online database to help you monitor toxicity alarms and generate data models;

UIntegrated solid phase extraction technology SPE (optional) device, the measured water samples can be concentrated and tested, and the concentration ratio can reach 1000 times.

UOther test probes and instruments (optional) can be integrated, such as the s::can uv-vis UV spectrophotometer, which can produce a more powerful early warning system for water quality.


Maintenance cycle

Because the new instrument iTOXcontrol is integratedFully automatic culture of bioluminescent bacteria, so it is possible to reduce the maintenance cycle to once every two weeks.At the same time, the consumption of luminescent bacteria and culture medium has also dropped to 50%.The luminescent bacteria stored in the second tank are also kept in fresh state and are automatically cultured after use in the first tank, which also ensures the effectiveness and sensitivity of the bacteria.

2.jpg


New iTOXcontrol related technical parameters and features:


UInstrument is more compact and smaller: 50 x 50 x 183 cm (D x W x H)

UCommunication: 4-20 mA signal output, CP/IP protocol

UOptional: Telephone communication module or add RS232 protocol communication

UThis is a very reliable and low-operation and low maintenance cost instrument

UTOXview software saves and evaluates data and operation results

UCan assemble automatic chlorine removal device

UOptional: Can be used in conjunction with test TOC, BOD, turbidity, SAC254, NO3-N and uv-vis probes


3.jpg




Technical parameters

ØTest water sample temperature range: 15-30 degrees Celsius.

ØOperating environment (ambient temperature):

General: 15-30 degrees Celsius;

Special environment: 30-40 degrees Celsius (requires air conditioning)

ØCommunication:

Standard configuration: TCP/IP protocol and 4-20 mA output

Optional: Add 56k USB module

Optional: The extended COM port is used by the MODBUS data communication protocol

ØInstrument cabinet: Protection level IP 31

ØWater sample dosage: + 10 ml/h (4, 5 ml/each test, 2 times/hour)

ØReference water sample dosage: + 100 ml/h each test, without chlorine, optional solenoid valve to control the reference water sample dosage

Assembled automatic chlorine removal device

ØWastewater discharge

Drainage pipe: Wastewater and control test waste liquid: 20 mm (external)


3.jpg

hardware

iTOXcontrol is assembled in an instrument cabinet with good ventilation:


UInstrument size: 50Small50Small110 Smart (DSmallWSmallH), lower small motorcycle height 73 cm

UWeight: + 90 kg (optional air conditioner: 20 kg)

UGlass front door, lockable to protect the instrument

UBuilt-in computer and keyboard and mouse touchpad, which can read and write DVD drives

U17-inch LCD screen, Ethernet interface, COM port and USB connection

UPower supply: 220V - 50 Hz or 110V - 60Hz.


software

TOXcontrol Engine software runs on Windows XP to control the settings and operation of the instrument.TOXview data storage and processing software is used to calculate and process related data and test results, as well as user settings for alarm limits.The status and status of the instrument are displayed on the status page of the TOXcontrol Engine software.The operation of the instrument iTOXcontrol is carried out in accordance with the instructions issued by the TOXcontro Engine software.Users can set the relevant parameters of the instrument to meet the different operating requirements of the user.The TOXview database will save the data used for the result calculation and the test results used.At the same time, the Toxview software will automatically generate charts related to the data used for users to view.




Collection