DB5301A DaqPro 8-channel collector
Date:2017-04-01 16:20:31

Instrument Introduction





DqPRO_DB5301A_01.jpg

dB5301AtypeDA go to pro 8The channel collector is a compact, high efficiency independent collector, and the measured values ​​can be displayed graphically on the host.The instrument can be set up alone or throughDA go to labSoftware for data download and advanced settings.











Instrument features





DqPRO_DB5301A_02.jpg

ØUp to accurate measurement16bit resolution8Input signals

ØAn alarm output

ØThe measured values ​​can be displayed graphically

ØMaximum sampling frequency can be reached4000Every second

ØEach sensor can be set separately

ØThe storage space is512KB RAM

ØRechargeable7.2VBattery-powered













Main technical parameters



DqPRO_DB5301A_03.jpg

ØInput Type

  • 4to20MAsensor

Range: 4 to 20mA

Resolution: 47.6μA

Accuracy: ±0.5%

Loop impedance: 21Ω

  • 0to50MVsensor

Range: 0 to 50mV

Resolution: 3μV

Accuracy: ±0.5%

  • 0to10Vsensor

Range: 0 to 10V

Resolution: 200μV

Accuracy: ±0.5%

  • NTCTemperature sensor

NTC: 10/100 KΩ resistor

Range: -25 to 150℃

Resolution: 0.05℃

Accuracy: ±0.5%

  • PT-100Temperature sensor

Range: -200 to 400℃

Resolution: 0.1℃ (7mΩ)

Accuracy: -200 to -50℃ ±0.5%; 50 to 400℃ ±0.5%; -50 to 50℃ ±0.5℃

  • JType thermocouple temperature sensor

Range: -200 to 1200℃

Resolution: 0.1℃

Accuracy: -200 to -50℃ ±0.5%; 50 to 1200℃ ±0.5%; -50 to 50℃ ±0.5℃

Cold junction temperature compensation: ±0.3℃

  • KType thermocouple temperature sensorRange: -250 to 1200℃

Resolution: 0.1℃

Accuracy: -250 to -50℃ ±0.5%; 50 to 1200℃ ±0.5%; -50 to 50℃ ±0.5℃

Cold junction temperature compensation: ±0.3℃

  • TType thermocouple temperature sensor

Range: -200 to 400℃

Resolution: 0.1℃

Accuracy: -200 to -50℃ ±0.5%; 50 to 400℃ ±0.5%; -50 to 50℃ ±0.5℃

Cold junction temperature compensation: ±0.3

  • Pulse counter (port only1)Range: 0 to 65000

Resolution: 1 count

Frequency range: 0 to 25Hz

Input signal: 0 to 5V

Input impedance: 470Ω

  • Frequency (port only1)

Range: 20Hz to 4KHz

Input signal: 0 to 5V

Input impedance: 470Ω

  • Built-in temperature sensor

Range: -25 to 70℃

Resolution: 0.1℃

Accuracy: ±0.5℃

  • GeneralAtoD

Noise: 30μV

Internal linear error: ±0.08% FSR

Calibration error: 0.1%

  • Open collector output

Maximum current source: 50mA (with fuse)

Maximum input voltage: 5V

Input impedance: 50Ω

ØHardware parameters

  • communicationUSB1.1

  • Instrument power supply (three options)

Built-in rechargeable 7.2V NiMH battery

Optional external 12V DC power input

  • Operating temperature0 to 50℃

  • Sampling settings

Memory: 512KB

Analog channel sampling rate: variable, 4000 per second to once an hour (for a single channel)

Analog sampling resolution: 16 bits

Channel spacing: 80dB

  • Host placement box

ABS clinker box

Size: 18.2 x 10 x 2.8 cm

Weight: 450g

  • Compatibility standardsCE\FCC













Origin:Fourtec, USA





Collection