Main functions
üCO, CO measurement by NDIR method or PID2, no, so2, sf6, THC
üMeasurement of NH by TDLS (Tunable Diode Laser)3, HCl and CH4Equivalent gas
üBy extending the principle of NDUV or electrochemical measurementmodule that can measure O2, O3or gases such as NO2
Application areas
This analyzer is especially suitable for industrial process gas analysis, process control and regulation, and laboratories.EasyCEM can measure up to 4 gases simultaneously as a single component or multi-component analyzer.Due to the special design, it can be adjusted according to the corresponding measurement.Many possible options ensure the use of analyzer EasyCEM in many applications.
Features
ü1-4 kinds of gases can be measured
üThe measurement range is expressed in ppm and Vol%
üRS232 interface
ü19-inch shell
üEasy to operate
How it works
üNDIR: Non-diffusion infrared analyzer
üNDUV: Non-diffusion UV analyzer
üTDLS: Laser Gas Analyzer
üPID: Photoionization detector
üEC: Electrochemical sensor
Main technical parameters
Measurement range (minimum, other ranges can be customized):
project | scope | Analyzer Type |
co2: | 0-40 ppm | n Dir |
no: | 0-1000ppm | n Dir |
no2: | 0-500 ppm | ND UV |
so2: | 0-100 ppm | n Dir |
co: | 0-500 ppm | n Dir |
THC: | 0-5% Vol | n Dir |
VOC: | 0-50 ppm or 0-5000 ppm | PID |
sf6: | 0-50 ppm | n Dir |
NH3: | 0-100 ppm | TD LS |
HCl: | 0-50 ppm | TD LS |
CH4: | 0-100 ppm | n Dir or TD LS |
O2: | 0-25% | EC |
Accuracy: ±2% of the measurement range
Repeatability: ±0.5 % (zero point); ±0.5 % (span range)
Signal response time: t90 < 2 s. (60 l/h); t90 < 1 s (180 l/h)
Operating temperature: 0-45℃
Temperature coefficient: < 1% range/10K
Warm-up time: < 3 minutes
Output signal type: digital model, RS232
Power supply: 230V/50Hz or 115V/60Hz
Size: 19 inches, 64cm
Weight: about 8-14kg
Optional: integrated pump, 1-10V DC analog signal output, alarm device, air intake filtration, flowmeter
Origin: MCZ, Germany